When the simplest box is the right box

Static Pass Box

A static pass box has no fan and no filter. In the right place that is not a shortcoming. It is the whole point. The skill is knowing exactly which place that is.

  • Same-grade transfers
  • Electromagnetic interlock
  • SS 304 / SS 316
  • No HEPA filter, by design

What a static pass box is

The simplest transfer hatch is the static pass box. It is an interlocked, stainless-steel chamber set into a cleanroom wall, with no air treatment inside. It lets material cross between two areas of the same cleanliness grade, without a person walking through.

  • ISO 14644-1
  • EU-GMP Annex 1
  • Schedule M
  • WHO-GMP
  • SS 304 / SS 316

The one-line test

Same grade on both sides of the wall? A static box is enough.

Change the grade and the same box becomes the cheapest way to move contamination across a boundary you paid to create.

You skip the cost, the power and the validation of a fan and filter. "Static" means the air in the box just sits there. The box isolates, it does not clean.

That makes it exactly right in one situation and quietly wrong in another, which is what choosing one comes down to. The rest of this page is about telling those two situations apart, because the box itself is the easy part.

Fabtech has delivered 2,000+ controlled environments since 2004, across every industry that needs controlled air. Each one was designed, built and validated in-house by one accountable team. So when we tell you a static box is enough, it is the same team that would have sold you the dynamic one saying so.

Air treatment
None. No HEPA filter and no blower, by design
Suits
Same-grade transfers, low risk
Interlock
Electromagnetic, both doors can never open together
Body
SS 304 as standard, SS 316 for aseptic or corrosive duty
Typical step across a grade boundary
15 Pa
States in one transfer: load, wait, remove
3
Standard chamber, 2 ft cube to 3 ft cube
600 to 900 mm
Germicidal lamp, fitted as standard
UV
The decision

When a static box is right, and when it's a false economy

A static box earns its place by doing less. The same box, in two different walls, is either the correct specification or the cheapest possible mistake. Here are both, drawn side by side.

Plan view, two wallsSchematic, not to scale
The same static pass box in two walls, stacked. Above, both rooms are ISO Class 7 and the chamber stays as clean as the rooms. Below, an unclassified corridor opens into an ISO Class 7 suite across a 15 Pa step and the particle load enters the chamber with the material.STATE A / SAME GRADEBoth sides at the same gradeGRADE LEVELNO STEPISO CLASS 7ISO CLASS 7DOOR ADOOR BBoth sides ISO Class 7, no stepChamber air is the same airNo airflow to qualify, no filter to testNothing to purge. Isolation is the job.Correct specificationSTATIC PASS BOXSTATE B / ACROSS A BOUNDARYUnclassified corridor into a suiteGRADE LEVEL+15 Pa STEPUNCLASSIFIED CORRIDORISO CLASS 7DOOR ADOOR BCorridor air is unclassifiedParticles enter with the materialNo fan and no filter is fittedNothing purges it. Specify dynamic.False economySPECIFY A DYNAMIC PASS BOX
  • Material route
  • Unfiltered air and its particle load
  • Built structure

FIG. 01One chamber, two walls. State A is a same-grade transfer, where a static box is the correct specification. State B is the same box across a 15 Pa grade boundary, where unfiltered air and its particle load enter the chamber and nothing is fitted to remove them.

A static box earns its place by doing less. Use one when

  1. 01

    Both sides are at the same grade

    And the transfer is low-risk: moving documents, tools or packaged items within a classified zone, or between two areas held to the same standard.

  2. 02

    A dynamic box would be over-specified

    Paying for a fan, a filter and HEPA-integrity validation that the boundary simply doesn't call for.

It becomes a false economy the moment the material has to step up a cleanliness grade. Picture an unclassified corridor opening into a classified suite. A static box does nothing to the particles the material carries from the dirtier side. It just moves the contamination across with it.

Dynamic pass box, HEPA airflow VHP pass box, bio-decontamination

State A

Same grade both sides

Static pass box, correctly specified

State B

Across a grade boundary

Static pass box, wrongly specified

Pressure differential across the wall
None, same grade
About 15 Pa
Air entering the chamber
Advantage. Same grade air
Unfiltered air
Particle load carried in
Advantage. None to speak of
Yes, with the material
What purges the chamber
Nothing needs to
Nothing is fitted
The right specification
Advantage. Static pass box
Dynamic pass box
Validation that applies
Interlock, cleanability, UV
Adds airflow and HEPA integrity

FIG. 02The same chamber, judged against the same six questions, in the two walls drawn above.

A static box is the right call when both sides sit at the same cleanliness grade. Once the grade changes, the box would carry particles across the boundary with the material. A dynamic box is then usually the safer specification.

Where a static box fits, and where it doesn't, is a judgement best made when the room's grade boundaries are designed, not picked from a catalogue.

Which box

Static, or step up?

Read the wall, not the catalogue. The boundary decides the box, and there are only three answers.

Which pass box suits which cleanliness boundary
The boundaryThe boxTHIS PAGEWhat the box has to do
Both sides at the same grade, low riskStatic pass box, this pageIsolate the two rooms during transfer. Nothing more is asked of it.
Any step up in cleanliness classDynamic pass boxStrip particles off the material with HEPA-filtered airflow before the clean door opens.
Into a sterile or aseptic areaVHP pass boxBio-decontaminate the chamber and its load, not just remove particles.

FIG. 03Boundary-to-equipment mapping, as specified on Fabtech projects.

The full side-by-side detail, air treatment, validation depth and cost, is set out in the static vs dynamic vs VHP comparison. Static vs dynamic vs VHP

Working principle

How a static pass box works

The static pass box working principle is deliberately plain, and the plainness is the feature.

No fan. No filter. One active idea:the electromagnetic interlock.

Plan view, three statesInterlock shown by leaf state
  1. Static pass box, state 01: Load and close. Door A open, door B locked.DOOR ADOOR BOPENLOCKEDINTERLOCK HOLDS DOOR B

    01

    Load and close

    An operator places the material inside and closes the door on their side. The electromagnetic interlock locks the opposite door, so both can never be open at once.

  2. Static pass box, state 02: Wait. Door A closed, door B closed.UVDOOR ADOOR BCLOSEDCLOSEDUV LAMP FITTED, EMPTY CHAMBER

    02

    Wait

    The material sits in the sealed chamber. A UV germicidal lamp is fitted as standard. It can be run for limited line-of-sight surface disinfection of the empty chamber between transfers. It treats directly-lit surfaces only, not shadowed or packaged material.

  3. Static pass box, state 03: Remove. Door A locked, door B open.DOOR ADOOR BLOCKEDOPENINTERLOCK HOLDS DOOR A

    03

    Remove

    A colleague on the other side opens their door and removes the material. The first door is locked for as long as the second one is open.

  • Door free to open
  • Chamber and wall

FIG. 04The three states of a static transfer. The electromagnetic interlock is the only active component: it is what guarantees the two rooms are never joined by an open air path.

There is no HEPA filter and no blower, and that is deliberate. A static box is not trying to clean the material. It is trying to isolate the two rooms. A clean door never has to open onto the corridor, and nobody has to carry contamination across on foot.

When both sides are already at the same grade, that is enough.

Anatomy

Key components of a static pass box

There is not much inside a static box, and that is the design. Two doors that cannot open together, a chamber you can clean properly, and a lamp. Nothing else has to keep working for the box to do its job.

01

Electromagnetically interlocked doors

The core feature: both doors can never open together, so the two rooms are never joined by an open path.

02

Stainless-steel body

SS 304 as standard, and SS 316 where the duty is aseptic or the cleaning chemistry is corrosive.

03

Coved internal corners

So there is no square corner for residue to sit in, and the chamber can be wiped out properly.

04

UV germicidal lamp

Fitted as standard: a between-transfers treatment for the empty chamber, limited to the surfaces the lamp can actually see.

05

Door interlock indicators

Status lights that tell an operator which door may open, so the interlock is legible from both sides of the wall.

06

No HEPA filter and no blower

By design. That absence is what makes it a static box, and it is why the qualification that follows it is shorter.

Standard chambers run from 600 × 600 × 600 mm (2 ft cube) up to 900 × 900 × 900 mm (3 ft cube). Within that, the opening and the depth follow the biggest item you need to pass, and the wall the box sits in.

Where they're used

Where static pass boxes are used

Static boxes suit transfers within a single classified area, or between two areas of the same grade.

Where the cleanliness grade changes across the wall, a dynamic or VHP box replaces it. That is the only test worth applying, and it is worth applying it wall by wall rather than once for the whole building.

Standards & validation

Standards and validation

Whether a static box is allowed on a given wall is not really a question about the box. It is a question about the two rooms it joins.

The grades on either side decide it, and those grades are the ones set by ISO 14644-1. For pharmaceutical manufacture, the same expectations reach you through EU-GMP Annex 1. In India they also arrive through Schedule M and WHO-GMP. None of that is written about static boxes. It is written about the rooms either side of one.

The good news is the paperwork. Qualifying a static box is genuinely less work than qualifying a dynamic one. There is no airflow to qualify and no filter to integrity-test. What is left is short, and it is set out here.

What gets qualified on a static pass box, and what does not
What gets qualified on a static pass boxTHIS PAGEWhat that means
Interlock functionBoth doors must never be open at once.
Construction and cleanabilityThe body and its coved corners can be cleaned properly.
UV functionThe lamp is fitted as standard, so it is checked.
HEPA integrity testing (PAO/DOP)Not run. No filter is fitted to test.
Airflow qualificationNot run. No fan, so there is no airflow to qualify.
On a dynamic boxThose last two sit inside its DQ/IQ/OQ/PQ.

FIG. 05Qualification scope for a static transfer hatch, against ISO 14644-1 and EU-GMP Annex 1 room grades.

That shorter qualification is a real part of why a static box costs less to own, not just less to buy. The paperwork is lighter, so there is less for your validation team to carry. We handle the equipment validation and AMC for the unit either way, so the shorter route doesn't land back on your desk.

Cleanroom validation and qualification ISO classification of cleanrooms

How Fabtech approaches it

How Fabtech approaches static transfer

Tell us the two rooms a box would sit between, their grades, and what's crossing the wall. Our engineers will give you a straight answer on whether a static box is the right economy, or whether that boundary needs more. If it helps the call, we can walk you through a named static-pass-box reference too.

Fabtech Cleanrooms is a cleanroom equipment manufacturer that designs, builds and validates controlled environments across India. It manufactures static pass boxes in-house, supplied both as standalone products and integrated within turnkey cleanroom projects.

Deciding where a static box is the right economy is part of that design work. Choosing a static box rather than a dynamic one is a grade-boundary decision. Get it right and you are neither paying for airflow you don't need, nor under-specifying a boundary that does.

Fabtech has delivered HEPA-filtered, classified cleanrooms for pharma and research clients across India. That work includes Ajanta Pharma, Desano and German Remedies, plus BSL-3 facilities at PGICH Noida and ASMC Shahjahanpur. It is the same contamination-control engineering that decides where a static box is and isn't acceptable.

Turnkey cleanroom solutions Cleanroom equipment range Talk to an engineer about your transfer points

Capability record
FABCLEAN / MATERIAL TRANSFER
  • Ajanta PharmaOSD facility, Aurangabad
  • DesanoPharma cleanrooms, Bangalore
  • PGICH NoidaBSL-3 facility
  • ASMC ShahjahanpurBSL-3 facility
FAQ

Static pass boxes, common questions

A static pass box is an interlocked transfer chamber with no air treatment inside. It moves material between two areas of the same cleanliness grade, without a person walking through. It isolates the two rooms. It does not clean the material.